Abstract Magnetic nanoparticles have garnered significant attention for various applications in the high-kHz-to-MHz range, although their magnetic characterization at these operational conditions has been limited. However, a number of recent studies have showcased high-frequency and high-field amplitude AC magnetometry loop tracers capable of retrieving the magnetic AC hysteresis curve and the associated magnetic properties. In this paper, we present an easily constructable loop tracer that is retrofitted into an existing AC calorimetry setup. This enables the loop tracer to function simultaneously with the AC calorimetry setup and also to be run as a high-frequency AC susceptometer. The loop tracer is shown to work in the frequency range 160–922 kHz with maximum applied fields from 18 to 46 mT depending on the frequency. An iron oxide nanoflower sample is used to test the loop tracer, showcasing high reproducibility in measured magnetic parameters as well as quantitative agreement between the different measurement methods in the setup.

Fields of Study & Subjects

Characterization and Applications of Magnetic Nanoparticles Magnetic Properties and Synthesis of Ferrites Geomagnetism and Paleomagnetism Studies

Publication details

Journal / source
Measurement Science and Technology
Volume
37
Issue
11
Pages
115901–115901
Publication date
6 February 2026