Experimental and Monte Carlo evaluation of a GAGG-BGO phoswich detector using 3D-printed Tc-99m FLUX phantoms

Read full text Summarise with VIRAT
Authors: Nuray Yavuzkanat, Vakkas Bozkurt, Taylan Yetkin, Sezgin Aydin, Sefa Ertürk
Cited by: 0
References: 29

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

Abstract This study presents an experimental and Monte Carlo-based evaluation of a compact GAGG(Ce)-BGO phoswich detector for detector-level Tc-99m imaging. Single-projection measurements obtained with a 4 × 4 detector module demonstrated that simple geometric structures from a 3D-printed phantom can be clearly resolved under collimated acquisition. Experimental images show good qualitative agreement with GATE-based simulations, confirming that the observed image features are primarily governed by detector configuration and source geometry. Spatial resolution analysis yields an intrinsic resolution of 4.3 mm from simulated point-source PSF/LSF and an effective image-based resolution of 5.9 mm derived from ESF/LSF analysis, reflecting additional blurring contributions inherent to image-based measurements. Geant4 simulations indicate a GAGG-to-BGO light output ratio of approximately five, consistent with experimental characterization and GATE-based image mapping. Compared to a reference NaI(Tl) detector, the phoswich configuration exhibits reduced background noise and improved signal-to-noise and contrast-to-noise ratios, highlighting its potential for scalable SPECT detector applications. The validated experimental–simulation framework provides a basis for future detector performance optimization, including depth-of-interaction–assisted image reconstruction and image quality enhancement through alternative collimator geometries and materials.

Fields of Study & Subjects

Radiation Detection and Scintillator Technologies Medical Imaging Techniques and Applications Cardiac Imaging and Diagnostics

Publication details

Journal / source
Physica Scripta
Volume
101
Issue
12
Pages
125303–125303
Publication date
23 March 2026