GRANITE: high-resolution imaging and electrical qualification of large-area TPC electrodes

Read full text
Authors: Shumit Mitra, Alexander Deisting, Jan Lommler, Uwe Oberlack, Fabian Piermaier, Quirin Weitzel, D. Wenz
Cited by: 1
References: 26

Abstract

Abstract Next-generation dual-phase time projection chambers (TPCs) for rare event searches will require large-scale, high-precision electrodes. To meet the stringent requirements for high-voltage performance of such an experiment, we have developed a scanning setup for comprehensive electrode quality assurance. The system is built around the GRANITE (Granular Robotic Assay for Novel Integrated TPC Electrodes) facility: a gantry robot on top of a 2.5 m× 1.8 m granite table, equipped with a suite of non-contact metrology devices. We developed a coaxial wire scanning head to measure and correlate localized high-voltage discharge currents in air with high-resolution surface images. We find that the identified discharge `hotspots' are transient and show no significant correlation with static visual features. Next, we established a quantitative relationship between artificially induced abrasive surface damage on the wires and a reduction in the discharge inception voltage. This work provides a novel non-invasive tool for qualifying wires dedicated for use in electrodes for future low-background experiments.

Fields of Study & Subjects

Particle Detector Development and Performance Neutrino Physics Research Radiation Detection and Scintillator Technologies

Publication details

Journal / source
Journal of Instrumentation
Publisher
IOP Publishing
Volume
21
Issue
01
Pages
P01034–P01034
Publication date
1 January 2026