Design and evaluation of front-end electronics prototype for the HIAF-HIRIBL gamma array detector

View at publisher
Authors: Longjie Wang, Xianqin Li, Zengxiang Wang, Duo Yan, Wei Zhang, Shitao Wang, Yuhong Yu, X. Niu, Haibo Yang
Cited by: 0
References: 12

Abstract

Abstract The Institute of Modern Physics, Chinese Academy of Sciences, is currently constructing the High Intensity Heavy Ion Accelerator Facility (HIAF). HIAF is capable of accelerating heavy ions ranging from protons to uranium, delivering high-energy ion beams with intensities of up to 10 11 ions per second. The High-Rigidity Radioactive Ion Beam Line (HIRIBL) is a key component of HIAF. A high-performance GeV-energy muon beam can be generated using the HIAF-HIRIBL facility. The gamma array detector is a critical component of HIRIBL. The gamma array detector utilizes a thallium-doped cesium iodide (CsI(TI)) crystals coupled with avalanche photodiodes (APDs), capable of detecting gamma rays as well as other types of charged particles. The detector array consists of multiple detection units. A group of 32 detection units is connected to the readout electronics, providing 32 independent readout channels. The readout electronics system comprises a front-end amplification unit (FAU), which integrates 32 charge-sensitive amplifier modules (CSAs), and a readout control unit (RCU). Additionally, a high-voltage power supply is employed to provide the bias voltage to the crystal detector. The charge-sensitive amplifier features adjustable gain, enabling the processing of signals with both high and low amplitudes. The signal enters the readout control unit, where it is processed by the ADC and FPGA, and the resulting data is transmitted to the data acquisition system via a 6 GSPS fiber-optic link. The prototype's readout electronics system has been manufactured and subjected to comprehensive electrical performance testing, including baseline, noise, and channel linearity tests. The readout electronics system was tested in conjunction with a cesium iodide detector. When tested with a 60 Co radioactive source, the results indicated that the system's dual-peak energy resolutions reached 7.39% and 6.52%, respectively.

Fields of Study & Subjects

Particle Detector Development and Performance Radiation Detection and Scintillator Technologies Advanced Semiconductor Detectors and Materials

Publication details

Journal / source
Journal of Instrumentation
Volume
21
Issue
02
Pages
T02010–T02010
Publication date
1 February 2026
Language
EN