Response modelling of a-Si:H detector read by TetrAMM picoammeter exposed to a clinical electron beam

Read full text
Authors: F. Cittadini, K. Kanxheri, M. Ponzecchi, C. Talamonti, M. Menichelli, R. Sacchi, S. Giordanengo, A. Vignati, U. Deut, D.M. Montalvan Olivares, S. Aziz, L. Barraud
+53 more A. Bashiri, C. Buti, L. Calcagnile, D. Calvo, D. Caputo, A.P. Caricato, R. Catalano, M. Cazzanelli, G. de Cesare, R. Cirio, G.A.P. Cirrone, T. Croci, G. Cuttone, S. Dunand, M. Fabi, L. Frontini, C. Grimani, M. Guarrera, H. Hasnaoui, G. Insero, M. Ionica, G. Konstantinou, F. Lenta, V. Liberali, N. Lovecchio, M. Martino, G. Maruccio, G. Mazza, A.G. Monteduro, A. Morozzi, A. Nascetti, S. Pallotta, D. Passeri, M. Pedio, M. Petasecca, G. Petringa, F. Peverini, P. Placidi, M. Polo, A. Quaranta, G. Quarta, P. De Remigis, S. Rizzato, F. Sabbatini, R. Sandhu, A. Stabile, Tosti L., M.M.S. Vasquez, A. Verdini, M. Villani, R.J. Wheadon, N. Wyrsch, L. Servoli
Cited by: 1
References: 10

Abstract

Abstract The INFN HASPIDE project investigates hydrogenated amorphous silicon (a-Si:H) as an innovative detector technology for dosimetry in FLASH radiotherapy. Conventional dosimeters, such as ionization chambers, show non-linear behavior when exposed to ultra-high dose-per-pulse beams used in FLASH radiotherapy, where several Grays are delivered in microsecond pulses. Thanks to its radiation hardness, mechanical flexibility and compatibility with large-area deposition, a-Si:H is a promising material for in-vivo and surface dosimetry. This work presents the response of a thin a-Si:H detector exposed to a high-dose-per-pulse 10 MeV electron beam, together with an analytical model used to describe the single-pulse dynamics and extract quantitative parameters. Measurements were performed at the University of Torino using a modified clinical LINAC operated in Ultra High Dose Rate regime, where the single-pulses satisfy the FLASH criteria. The a-Si:H detector, read out with a TetrAMM picoammeter (100 kHz sampling frequency), resolves individual radiation pulses with high repeatability. Both pulse charge and peak current scale linearly with the applied electric field up to 6 V (2.4 V/μm), beyond which the readout electronics saturates. The analytical model is validated through direct comparison with raw data, showing excellent agreement in extracted pulse parameters.

Fields of Study & Subjects

Radiation Therapy and Dosimetry Advanced Radiotherapy Techniques Particle Detector Development and Performance

Publication details

Journal / source
Journal of Instrumentation
Publisher
IOP Publishing
Volume
21
Issue
03
Pages
C03005–C03005
Publication date
1 March 2026
Language
EN