MPGD with a 3D-printed Thick-GEM as sole gain element

Read full text
Authors: J. Collins, M. Hohlmann
Cited by: 0
References: 28

Abstract

Abstract We present the first micropattern gaseous detector that employs a small 3D-printed Thick-GEM as its sole gain element. The detector produces detectable avalanche signals under the tested conditions without the need for pre-amplification by additional gain elements. We describe the design, quality control, assembly, and test of this detector. The 10 cm× 10 cm active area of the Thick-GEM features three separate sectors with different sizes of the clearance rim annuli (0.1 mm, 0.15 mm, 0.2 mm) around the 0.7 mm diameter holes. The gas gain is found to depend strongly on the rim size. When operated in Ar/CO 2 70:30 gas, the sector with 0.15 mm annulus rims reaches a gain above 10 4 while operating in a stable manner with an acceptably low discharge rate. The gain reach under stable operation is found to be considerably lower in the other two sectors. The gas gain shows a characteristic time-dependence as it rises quickly in the first hour of operation and then drops slowly over the next 24 hours and subsequently stabilizes.

Fields of Study & Subjects

Particle Detector Development and Performance CCD and CMOS Imaging Sensors Analog and Mixed-Signal Circuit Design

Publication details

Journal / source
Journal of Instrumentation
Publisher
IOP Publishing
Volume
21
Issue
08
Pages
P08027–P08027
Publication date
1 August 2026