WaveAtom based denoising approach for enhancing non-invasive fetal phonocardiography signal analysis

View at publisher
Authors: Manju T. Kurian, P Rajalakshmy
Cited by: 1
References: 26

Abstract

The non-invasive acoustic recording of fetal cardiac activity, known as fetal phonocardiogram (fPCG), provides valuable information for the early identification of fetal cardiac anomalies. However, dynamic artifacts, background noise, and powerline disturbances usually degrade signal quality, challenging analysis. The article proposes a multiscale frequency-time analysis denoising architecture. Robust peak detection is possible using a statistical envelope based on the Hilbert transform, following moving-average smoothing. Applying predefined rate intervals, physiological gating differentiates maternal and fetal heart rate (HR) streams. Fetal and maternal peaks can be accurately distinguished using only their physiological rate ranges, as this involves gating. Metrics such as the correlation coefficient, Root Mean Square Error(RMSE), and Signal-to-Noise Ratio (SNR) were used to evaluate system performance. The proposed method works well, especially in medical signal augmentation situations, where fine-grained oscillatory features are difficult to handle with conventional methods.

Fields of Study & Subjects

Phonocardiography and Auscultation Techniques Neonatal and fetal brain pathology ECG Monitoring and Analysis

Publication details

Journal / source
Biomedical Physics & Engineering Express
Publisher
IOP Publishing
Volume
12
Issue
2
Pages
025024–025024
Publication date
16 February 2026
Language