Verification and experimental validation of neutral atom beam source produced by L-PBF
Abstract
Abstract We report validation tests of a calcium atomic-beam source fabricated via Laser Powder Bed Fusion (L-PBF). The surface quality and elemental composition of the printed component were quantitatively assessed, allowing us to establish reference parameters for reliable operation in an ultra-high-vacuum environment. Safe operating conditions of the atomic oven were determined through a combination of simulations and experimental measurements. The ability of the device to deliver an atomic beam to the main experimental region—the electron/ion trap—was verified using atomic fluorescence imaging. Fluorescence spectroscopy was further employed to characterize the beam divergence, yielding an emission-cone half-angle of approximately 19° for atoms near the beam axis. A current of atoms on the order of 10 8 s −1 was estimated in the electron-trapping region, which is more than sufficient for anticipated electron-trapping and ion-trapping experiments.
Fields of Study & Subjects
Atomic and Molecular Physics
Laser Design and Applications
Laser-Plasma Interactions and Diagnostics
Publication details
- Journal / source
- Journal of Physics Communications
- Volume
- 10
- Issue
- 3
- Pages
- 031001–031001
- Publication date
- 1 March 2026
- Language
- EN