Diagnostic validation of a novel high-sensitivity cardiac troponin T assay

Read full text
Authors: Ben Völschow, Bianca Strauß, Nils Arne Sörensen, Niklas Thießen, Jonas Lehmacher, Betül Toprak, Beya Abdennadher, Farnaz Dehkordi, P M Haller, Lea Scharlemann, Alina Schock, Astrid Petersmann
+3 more Karen Friederike Gauß, Raphael Twerenbold, Johannes T Neumann
Cited by: 1
References: 29

Abstract

AIMS: High-sensitivity cardiac troponin (hs-cTn) assays are integral to evaluating suspected myocardial infarction (MI). A recently developed hs-cTnT assay requires diagnostic and prognostic evaluation. METHODS AND RESULTS: Patients presenting to the emergency department with suspected non-ST-elevation MI were prospectively evaluated. Discrimination was tested using the area under the curve (AUC). Diagnostic performance was assessed with previously derived cut-off concentrations and ESC 0/1 h algorithm. Head-to-head comparisons between the new and established assays were performed. Prognostic performance of hs-cTnT for 1-year mortality and major adverse cardiac events (MACE) was analysed using adjusted Cox regression. In 1415 patients, mean age 64 years and 64.2% male, MI was diagnosed in 12.1%. The novel TnT hs Gen 6 assay demonstrated excellent discrimination with an AUC on admission of 0.87 (95% CI: 0.84, 0.90) and 0.91 (95% CI: 0.89, 0.93) after 1 h. Using a 0 h cut-off

Fields of Study & Subjects

Acute Myocardial Infarction Research Coronary Interventions and Diagnostics Cardiac Imaging and Diagnostics

Publication details

Journal / source
European Heart Journal Acute Cardiovascular Care
Publisher
Oxford University Press (OUP)
Volume
15
Issue
7
Pages
535–544
Publication date
15 April 2026
Language
EN