One short circuit, two novel subcutaneous implantable cardiac defibrillator complications

Read full text
Authors: Alexander P. Bates, S Leighton, C Atha, Richard Balasubramaniam, Mark Sopher
Cited by: 0
References: 10

Abstract

Background: Short circuits from cardiac implantable electronic device (CIED) systems leak current to surrounding tissues, rendering them malfunctional and resulting in unpredictable behaviour. Case summary: A 74-year-old gentleman was admitted electively for subcutaneous implantable cardiac defibrillator (SICD) generator change. The generator was exchanged like-for-like without incident. During defibrillation testing (DFT), both shocks were unsuccessful, with the second shock causing a malfunction in the self-adhesive pads attached to the external defibrillator. Following brief cardiopulmonary resuscitation, new pads were attached, resulting in successful external defibrillation. Post-procedure analysis revealed the impedance within the device system was