Long-term safety and performance of an implantable IVC sensor for congestion-guided management in heart failure: 12-month results from the FUTURE-HF trial portfolio

Read full text
Authors: Kevin Damman, Nir Uriel, Mandeep R Mehra, Jeffrey Testani, M F Marat Fudim, Kunjan Bhatt, Rami Kahwash, Trejeeve Martyn, Scott Feitell, George Khabeishvili, Petr Neužil, Gabriel Sayer
+13 more Irakli Gogorishvili, Chris Critoph, Faisal Sharif, Ian Loke, RS Gardner, Ondřej Toman, Andrew Flett, Jasper J Brugts, Ali Vazir, Teresa Buxó, Alastair Gray, Barry R. Greene, PR Kalra
Cited by: 2
References: 23

Abstract

AIMS: Congestion signals heart failure progression and drives decompensation. Reliable management strategies remain poorly developed. We evaluated 12-months of congestion-guided clinical management following implantation of an inferior vena cava (IVC) sensor. METHODS AND RESULTS: Data were combined from two prospective studies (FUTURE-HF and FUTURE-HFII) (N=65, mean age 65.7±9.5 years; 75.4% NYHA III; 90.8% HFrEF). Patients recorded daily IVC parameters. Adjudicated safety outcomes, sensor-derived IVC area measurement versus CT imaging, medication adjustments, and clinical outcomes at 12-months were analysed.No adjudicated device or procedure-related serious adverse events occurred. Excellent correlation was observed between sensor-derived and CT-derived IVC area (n=44; R2=0.97; mean relative error