Do Filters Affect Humidity Levels Within Helmet-CPAP Powered by the Venturi System? Insights from a Bench-top Study
Abstract
OBJECTIVE: Helmet continuous positive airway pressure (H-CPAP) systems powered by Venturi devices are widely used for noninvasive respiratory support in acute hypoxaemic respiratory failure. Although mechanical filters are often added to reduce noise and improve environmental safety, their impact on gas humidification, flow, and FiO₂ delivery remains unclear. This bench-top study aimed to evaluate how the use of different filter types and configurations affects the humidity within the helmet. The secondary objective was to assess the changes in FiO₂ and flow associated with filter placement. METHODS: A bench-top study was conducted at the Respiratory Physiology Laboratory of the Italian University Hospital using a medium-sized CPAP helmet connected to a Venturi flow generator. Seven mechanical filters (5 HEPA and 2 HME) were tested in 3 configurations: no filter, 1 filter between the Venturi flowmeter and helmet gas-in, and 2 filters (post-flowmeter and Venturi air inlet). Temperature, relative humidity, flow, and FiO₂ were measured under 3 FiO₂ settings (0.35, 0.55, 0.75). RESULTS: The addition of filters led to a statistically significant reduction in the absolute humidity inside the helmet (P
Fields of Study & Subjects
Respiratory Support and Mechanisms
Neonatal Respiratory Health Research
Inhalation and Respiratory Drug Delivery
Publication details
- Journal / source
- Dimensions of Critical Care Nursing
- Publisher
- Ovid Technologies (Wolters Kluwer Health)
- Volume
- 45
- Issue
- 3
- Pages
- 149–154
- Publication date
- 1 April 2026
- Language
- EN