Implantable Tibial Neuromodulation Therapy Improves Symptoms of Urge Urinary Incontinence From the TITAN 2 Pivotal Study
Abstract
PURPOSE: The Medtronic implantable tibial neuromodulation system is a leadless, rechargeable, minimally invasive option intended to treat the symptoms of urge urinary incontinence (UUI), a common condition that significantly impairs quality of life. TITAN 2 is a pivotal study evaluating safety and effectiveness of the Medtronic implantable tibial neuromodulation system. The results through 12 months are reported. MATERIALS AND METHODS: This prospective, multicenter, single-arm investigational device exemption study enrolled eligible overactive bladder subjects with UUI. The primary end point was the percentage of subjects who had ≥ 50% reduction in UUI episodes/day after 6 months. Safety was characterized as the rate of adverse device effects (ADEs). Data are reported as mean ± SD or with 95% CIs. RESULTS: < .0001) at 6 months; at 12 months, 61% had ≥ 50% improvement. Change from baseline was statistically significant at 6 months for UUI episodes/day, voids/day (in urinary frequency subjects), and Urgency Perception Score, with improvements in all 3 symptoms maintained through 12 months. Twenty percent of subjects experienced ADEs through 12 months; 1 ADE was serious and was recovered/resolved. CONCLUSIONS: TITAN 2 met the primary and all secondary objectives by showing a statistically significant improvement of UUI, urinary frequency, and urgency symptoms. Clinical benefits and a favorable safety profile were maintained through 12 months.
Fields of Study & Subjects
Urinary Bladder and Prostate Research
Pain Management and Treatment
Neurological disorders and treatments
Publication details
- Journal / source
- The Journal of Urology
- Publisher
- Ovid Technologies (Wolters Kluwer Health)
- Volume
- 215
- Issue
- 6
- Pages
- 702–712
- Publication date
- 26 January 2026
- Language
- EN