Validation of the Hanvon KSY3600 Korotkoff-sound electronic blood pressure monitor in the general population according to the ISO 81060-2:2018

View at publisher
Authors: Yi-Xiang Wang, Yingjian Liu, Chongjie Chen
Cited by: 1
References: 3

Abstract

This study aims to evaluate the accuracy of the Hanvon KSY3600 Korotkoff-sound electronic blood pressure (BP) monitor in the adult population, in accordance with the ISO 81060-2:2018. A total of 90 participants were recruited for this clinical study. After excluding 3 individuals due to excessive BP variability, the remaining 87 participants were enrolled for the evaluation. All enrolled participants met the inclusion criteria, which specified ranges for arm circumference, SBP, and DBP. According to ISO 81060-2:2018 + Amd 1:2020, the mean ± SD of the differences between the test device and the reference BP measurements were -0.3 ± 2.35/-0.3 ± 1.72 mmHg (SBP/DBP), thus meeting the requirements of Criterion 1 (≤5 mmHg mean difference and ≤8 mmHg SD). The mean difference in SBP was -0.3 ± 1.11 mmHg, while that in DBP was -0.3 ± 0.90 mmHg; this also fulfilled the requirements of Standard 2, with standard deviations within the limits of ≤6.95 mmHg for both SBP and DBP. Both ISO criteria were therefore satisfied, showing that the Hanvon KSY3600 Korotkoff-sound electronic BP monitor conforms to the ISO 81060-2:2018 + Amd 1:2020. It can thus be recommended for BP measurements in adults, both in clinical settings and for home use.

Fields of Study & Subjects

Blood Pressure and Hypertension Studies Non-Invasive Vital Sign Monitoring Health Education and Validation

Publication details

Journal / source
Blood Pressure Monitoring
Publisher
Ovid Technologies (Wolters Kluwer Health)
Volume
31
Issue
4
Pages
193–197
Publication date
25 February 2026
Language
EN