Reply to “Comments on ‘Experimental Evaluation of Relationship Between Radiofrequency Heating Near Implanted Conductive Devices, Scanner-Reported B 1+rms , and Transmit Power’”

Read full text
Authors: David H. Gültekin, J. Thomas Vaughan, Devashish Shrivastava
Cited by: 1
References: 3

Abstract

We describe the relationship between the RF-induced temperature change in a conductive metallic implant during the MRI, B1+rms, and RF power for RF pulse sequences, RF coils, and Tx/Rx combinations. We show that both temperature and RF power change with RF pulse sequences, RF coils, and Tx/Rx combinations, whereas B1+rmschanges only with RF pulse sequences but not RF coils and Tx/Rx combinations. As a result, induced heating is associated with RF power propagation and loss corresponding to the B1+rms.