Development of a Quantum Blackbody Thermometer Toward Primary On-Orbit Thermometry

Read full text
Authors: Peter Beierle, Denis A. Tremblay, Noah Schlossberger, Christopher L. Holloway, Stephen Eckel, Eric B. Norrgard
Cited by: 0

Abstract

We present a roadmap to a deployable, intrinsically calibrated thermometer with long-term accuracy of 30 mK, exceeding existing on-orbit resistance-based thermometers. Our quantum blackbody thermometer is based on measuring fluorescence ratios of optically excited rubidium atoms in microfabricated vapor cells. The key advantage of the quantum blackbody thermometer is that long-term stability of the fluorescence ratios is guaranteed by the immutable physical properties (transition strengths) of the rubidium atom. This should be compared against resistance-based thermometers, such as platinum resistance thermometers, which may be calibrated with exceptional accuracy but are susceptible to temporal drift and shifts due to improper handling.

Fields of Study & Subjects

Calibration and Measurement Techniques Advanced Frequency and Time Standards Scientific Measurement and Uncertainty Evaluation

Publication details

Journal / source
IEEE Sensors Journal
Volume
26
Issue
19
Pages
28387–28392
Publication date
12 August 2026