A CSRR-IDC-Based Microwave Biosensor for Non-invasive Creatinine Detection

View at publisher
Authors: Teja Jetti, Andleeb Zahra, Swarnim Sinha, Syed Azeemuddin, Zia Abbas
Cited by: 0
References: 18

Abstract

Rapid detection of creatinine level is significant for the diagnosis of renal dysfunction. Conventional methods used for creatinine detection are not portable and cannot provide real-time results. This paper describes the design and validation of a microwave biosensor to overcome these limitations. The resonator is designed using a complementary split-ring resonator (CSRR) and an interdigitated capacitor (IDC) fabricated on a FR-4 substrate of dimensions 50×25mm2.Simulations showedaresonancefrequencyof2.013GHz, while measurement results yielded a resonance of 1.941GHz, with a deviation of 72 MHz. The fabricated device was tested with 10μL creatinine samples of 0.8 to 2.0mg/dL, which are in the physiolog ical range. A sensitivity of 0.77% is achieved for the detection of concentration change in chemical solutions. These results show the capability of the biosensor for real-time biomedical diagnosis.

Fields of Study & Subjects

Microwave and Dielectric Measurement Techniques Microwave Imaging and Scattering Analysis Biosensors and Analytical Detection

Publication details

Journal / source
IEEE Sensors Letters
Volume
10
Issue
4
Pages
1–4
Publication date
30 January 2026