Multimodal Microelectrode-Microcantilever Array for Electromechanical Analysis of Cardiomyocyte Tissue in Drug Testing

View at publisher
Authors: Haolan Sun, Longlong Li, Eung‐Sam Kim, Bong-Kee Lee, Dong‐Weon Lee
Cited by: 0
References: 14

Abstract

We present an integrated bio-monitoring platform combining a microelectrode array (MEA) with a silicon microcantilever array for simultaneous electrical and mechanical characterization of cardiomyocytes. Each cantilever features a micro-pillar with a diameter of$50 \mu \mathrm{m}$at its free end to detect contraction-induced deflection via a piezoresistive strain sensor, while 16-channel gold MEA ($50 \mu \mathrm{m}$electrodes, 10 kHz sampling) records field potentials. The cantilevers exhibit a sensitivity of 3.67$\text{mV} / \text{nN}$and a minimum detectable force of 160 nN. Compared with conventional AFM, the proposed platform enables parallel, alignment-free, and long-term measurements with improved throughput and signal stability, offering new possibilities for real-time studies of cardiac electromechanical coupling.

Fields of Study & Subjects

Advanced MEMS and NEMS Technologies Neuroscience and Neural Engineering Cellular Mechanics and Interactions

Publication details

Journal / source
IEEE International Conference on Micro Electro Mechanical Systems (MEMS)
Pages
1–4
Publication date
25 January 2026
Language
EN