Design, Modelling and Characterisation of a Miniature Fibre-Reinforced Soft Bending Actuator for Endoluminal Interventions

View at publisher
Authors: Xiangyi Tan, Aoife McDonald-Bowyer, Danail Stoyanov, Agostino Stilli
Cited by: 1
References: 16

Abstract

Miniaturised soft pneumatic actuators are crucial for robotic intervention within highly constrained anatomical pathways. This work presents the design and validation of a fibre-reinforced soft actuator at the centimetre scale for integration into an endoluminal robotic platform for natural-orifice interventional and diagnostic applications. A single-chamber geometry reinforced with embedded Kevlar fibre was designed to maximise curvature while preserving sealing integrity, fabricated using a multi-stage multi-stiffness silicone casting process, and validated against a high-fidelity Abaqus FEM using experimentally parametrised hyperelastic material models and embedded beam reinforcement. The semi-cylindrical actuator has an outer diameter of $18, \mathrm{~mm}$ and a length of $37.5, \mathrm{~mm}$. Single and double helix winding configurations, fibre pitch, and fibre density were investigated. The optimal 100 SH configuration achieved a bending angle of 202.9° experimentally and 297.6° in simulation, with structural robustness maintained up to $100, \mathrm{kPa}$ and radial expansion effectively constrained by the fibre reinforcement. Workspace evaluation confirmed suitability for integration into the target device envelope, demonstrating that fibre-reinforcement strategies can be effectively translated to the centimetre regime while retaining actuator performance.

Fields of Study & Subjects

Prosthetics and Rehabilitation Robotics Advanced Materials and Mechanics Soft Robotics and Applications

Publication details

Journal / source
2026 IEEE 9th International Conference on Soft Robotics (RoboSoft)
Publisher
IEEE
Pages
550–557
Publication date
7 April 2026