Transition-Edge Sensor Characteristics in a Modified Bias Circuit for Multiplexed Readout

View at publisher
Authors: S. J. Smith, Paul van der Hulst, Nicholas A. Wakeham, Edoardo Cucchetti, William Bertrand Doriese, Malcolm Durkin, Christian Kirsch, Joel N. Ullom, J. S. Adams, S. R. Bandler, James A. Chervenak, Tannaz Farrahi
+12 more Fred Michael Finkbeiner, Joshua D. Fuhrman, Samuel V. Hull, Richard L. Kelley, Caroline Anne Kilbourne, H. Muramatsu, Frederick Scott Porter, Kazuhiro Sakai, Ambarish C. Venkatasubraman, Edward J. Wassell, M. C. Witthoeft, Sang H. Yoon
Cited by: 0
References: 12

Abstract

The X-ray Integral Field Unit (X-IFU) on the NewAthena X-ray telescope will use a 1,504-pixel array of transition-edge sensor (TES) microcalorimeters, read out via time division multiplexing (TDM). To allow testing of multiplexer chips before TES integration, each unit cell includes a “TES-bypass” resistor that substitutes for a TES during characterization. After integration, this resistor remains as a high-resistance shunt: minimal for DC operation but can influence the frequency response due to its placement with the Nyquist inductor. We modeled the modified bias circuit, derived the small-signal response, and evaluated noise contributions, including Johnson noise from the bypass resistor. We analyzed the effects on pulse height, slew-rate, and energy resolution. Simulations were compared with measurements from prototype X-IFU microcalorimeters. Our results indicate that as long as the resistor value exceeds the TES dynamic resistance, performance is largely unaffected.

Fields of Study & Subjects

Superconducting and THz Device Technology Particle Detector Development and Performance Astronomy and Astrophysical Research

Publication details

Journal / source
IEEE Transactions on Applied Superconductivity
Volume
36
Issue
6
Pages
1–7
Publication date
12 February 2026
Language
EN