Noise Modeling and Calibration of a Two-Stage Cryogenic Charge Amplifier for the SPLENDOR Experiment

Read full text
Authors: J. Anczarski, Owen Andrews, T. Aralis, C. W. Fink, Noah Kurinsky, A. Phipps, A. Pradeep, Betty Young
Cited by: 0
References: 13

Abstract

The SPLENDOR Collaboration studies novel narrow-gap semiconductors and engineered a substrate agnostic detector platform to achieve$\mathcal {O}$(meV) energy sensitivity designed for low mass dark matter searches. This was achieved using low-capacitance and low-noise commercial CryoHEMTs in a split-stage topology integrated throughout a dilution refrigerator. Designed with a source-follow HEMT at the base temperature stage and a voltage amplifier at 4 K, this amplifier has input-limited voltage noise of 10$\text{nV}/\sqrt{\text{Hz}}$and current noise of100$\text{aA}/\sqrt{\text{Hz}}$at 1kHz. In agreement with this noise level and a photon calibration, this amplifier has a$\text{19} \pm \text{4}$electron resolution.

Fields of Study & Subjects

Neutrino Physics Research Astrophysics and Cosmic Phenomena Dark Matter and Cosmic Phenomena

Publication details

Journal / source
IEEE Transactions on Applied Superconductivity
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Volume
36
Issue
6
Pages
1800306–1800306
Publication date
13 April 2026
Language