Design and Evaluation of a Compliant Quasi-Direct Drive End-Effector for Safe Robotic Ultrasound Imaging

Read full text
Authors: Danyi Chen, Ravi Prakash, Vincent M. Wang, Zacharias Chen, Sarah Dias, Daniel M. Buckland, Leila J. Bridgeman, R. Siobhan, Brian P. Mann
Cited by: 0

Abstract

Robot-assisted ultrasound scanning promises to advance autonomous and accessible medical imaging. However, ensuring patient safety and compliant human-robot interaction during probe contact poses a significant challenge. Most existing systems either have high mechanical stiffness or trade performance for compliance. This paper presents a novel compliant end-effector designed to mount on robotic arms for safe and accurate robotic ultrasound imaging, using a quasi-direct drive actuator to achieve passive mechanical compliance and precise active force control. To evaluate the end-effector’s performance, we developed an ex vivo dynamic motion simulator platform for contact and scanning testing on tissue under simulated movements. The end-effector was evaluated against a UR3e robot arm using conventional force control strategies as a baseline. Single-point contact experiments from 2.5 N to 15 N show that the end-effector reduced force tracking RMS error by 80.1% on average. Trajectory scanning experiments at different speeds showed an average of 68.0% reduction in force tracking error. Statistically significant improvements were observed in four of six quantitative image quality and stability metrics using the end-effector. This work presents a novel approach for designing and evaluating compliant end-effectors, with the goal of improving safety and reliability in robotic ultrasound.

Fields of Study & Subjects

Piezoelectric Actuators and Control Ultrasound Imaging and Elastography Ultrasound and Hyperthermia Applications

Publication details

Journal / source
IEEE Transactions on Medical Robotics and Bionics
Volume
8
Issue
3
Pages
1383–1395
Publication date
1 June 2026
Language
EN