Introducing Timepix2-Lite: A Miniaturized Readout Interface Enabling Nanosecond-Scale Half-Life Measurement

Read full text
Authors: O. Pavlas, Benedikt Ludwig Bergmann, Michael Holik, Milan Malich, S. Pospı́šil, Petr Smolyanskiy, V. Vícha, Robert Filgas
Cited by: 0
References: 19

Abstract

This paper presents Timepix2 Lite, a compact readout interface for the Timepix2 hybrid pixel detector, designed to provide high spatial and temporal resolution in a wide range of radiation detection applications. The Timepix2 Lite enables simultaneous energy and precise timing measurements on the nanosecond scale and supports advanced real-time data acquisition and visualization through the TrackLab software framework. Its compact and flexible architecture, enabled by the miniaturized design, allows seamless integration into diverse experimental setups, including nuclear spectroscopy, stratospheric balloon missions, and nanosecond-scale half-life measurements. As a case study, the half-life of the second excited state in237Np was determined to be 67.5(7) ns for the 59.5 keV transition, consistent with previously reported results.

Fields of Study & Subjects

Particle Detector Development and Performance Radiation Detection and Scintillator Technologies Radioactive Decay and Measurement Techniques

Publication details

Journal / source
IEEE Transactions on Nuclear Science
Volume
73
Issue
6
Pages
2262–2270
Publication date
12 February 2026
Language
EN