Self-Operating Geometry Calibration for Cone-Beam Computed Tomography Using Brute-Force Search

View at publisher
Authors: Jiyong Shim, Woosung Kim, Hyunwoo Lim, Hyosung Cho
Cited by: 1
References: 23

Abstract

oAccurate scan geometry is essential for high-quality three-dimensional (3D) reconstruction in cone-beam computed tomography (CBCT). This study presents a practical and self-operating geometric calibration method that does not require dedicated calibration phantoms. It estimates three critical scan parameters—vertical shift (δv), horizontal shift (δu), and in-plane rotation angle (ϕ)—through a multi-stage brute-force search guided by mean structural similarity (MSSIM). Unlike conventional methods that rely on specially designed reference objects, the proposed approach operates solely on projection data of the object being scanned, ensuring greater flexibility and ease of use. The method was validated through simulations using a 3D Shepp-Logan phantom and experimental tests with an industrial micro-CT system and a cigarette lighter phantom. It effectively corrected geometric misalignments, enhancing image quality and reducing artifacts such as edge blurring and double contours. In conclusion, the proposed method enables accurate 3D reconstruction without requiring additional calibration setups, providing a reliable and convenient solution for practical CBCT imaging applications.

Fields of Study & Subjects

Medical Imaging Techniques and Applications Advanced X-ray and CT Imaging Digital Image Processing Techniques

Publication details

Journal / source
IEEE Transactions on Nuclear Science
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Volume
73
Issue
5
Pages
1791–1799
Publication date
27 February 2026