Characterization of Electrical Power Instabilities for High-Intensity Pulsed Therapeutic Ultrasound

Read full text
Authors: Virgil Accary, Thomas Payen, Alexandre Baudouard, Maxime Lafond, Cyril Lafon
Cited by: 0
References: 18

Abstract

Therapeutic ultrasound has been the focus of numerous recent investigations. Recent developments, such as histotripsy techniques, have led to greater electrical solicitation of ultrasound setups with high-intensity and short bursts, which can induce potential risks for patients and operators, as well as suboptimal performances. Neither acoustic nor electrical engineering standards are designed for analyzing such signals. Ultrasound setups are usually characterized by voltage-only (VO) and wattmeter power measurements under long-duration (>1 s) and low-power (

Fields of Study & Subjects

Ultrasound and Hyperthermia Applications Ultrasonics and Acoustic Wave Propagation Quality and Safety in Healthcare

Publication details

Journal / source
IEEE Transactions on Ultrasonics
Volume
73
Issue
3
Pages
296–304
Publication date
17 February 2026