Retention of a Detached Robotic Scissors Tip Cover in the Abdominal Wall: A Case Report

Read full text
Authors: Masatsugu Kojima, Toru Miyake, Soichiro Tani, Keiji Muramoto, Yusuke Nishina, Sachiko Kaida, Katsushi Takebayashi, Hiromitsu Maehira, Reiko Otake, Haruki Mori, Nobuhito Nitta, Miyuki Kimura
+3 more N. Ushio, Tomoharu Shimizu, Masaji Tani
Cited by: 0
References: 9

Abstract

Robotic surgery has become increasingly widespread; however, device-related complications specific to robotic platforms are rarely reported. We describe a case of rectal cancer treated with robotic-assisted abdominoperineal resection using the da Vinci Surgical System. At the end of the procedure, the tip cover of the robotic scissors detached and was inadvertently retained within the abdominal wall. It was invisible on plain radiography but was detected on postoperative computed tomography as a cylindrical structure beneath the rectus abdominis muscle. The patient underwent reoperation to retrieve the tip cover and recovered uneventfully, being discharged without further complications. This case highlights a rare but important complication of robotic surgery-detachment and retention of the tip cover of robotic scissors in the abdominal wall. Because tip covers may not be reliably detected on plain radiography, computed tomography is crucial for their identification. Strict counting protocols and heightened awareness are essential to prevent such events.

Fields of Study & Subjects

Minimally Invasive Surgical Techniques Hemostasis and retained surgical items Surgical Simulation and Training

Publication details

Journal / source
Asian Journal of Endoscopic Surgery
Volume
19
Issue
1
Pages
e70242–e70242
Publication date
1 January 2026