Acute Atrial Pacing Threshold Elevation in an Auricular Leadless Pacemaker During Monopolar Pulsed Field Ablation of the Right Inferior Pulmonary Vein: A First‐in‐Human Case

Read full text
Authors: Jorge O. Díaz, U Siddiqui, Carlos E. Arenas
Cited by: 2
References: 3

Abstract

INTRODUCTION: Leadless atrial pacemakers are increasingly used, yet interactions with pulsed field ablation (PFA) remain poorly described. METHODS AND RESULTS: We report the first documented case of transient loss of capture and pacing threshold elevation in an AVEIR leadless pacemaker temporally associated with monopolar PFA during right inferior pulmonary vein ablation. Threshold increased from 0.75 V @ 0.2 ms to 1.25 V @ 0.4 ms, recovering within 24 h. CONCLUSION: Monopolar PFA can transiently interfere with atrial leadless pacing, emphasizing the need for device monitoring during ablation.

Fields of Study & Subjects

Cardiac Arrhythmias and Treatments Cardiac pacing and defibrillation studies Atrial Fibrillation Management and Outcomes