Tether Entrapment and Fracture Preventing Release and Redocking of a Helix‐Fixation Atrial Leadless Pacemaker: Successful Resolution With a Protective Sleeve–Assisted Bailout Maneuver

Read full text
Authors: Shuhei Arai, Takamasa Ishikawa, Rimpei Ueno, Yosuke KAI, Haruka Miyazaki, Saori Chino, Kosuke Yoshikawa, Yuya Nakamura, Taku Asano, Toshiro Shinke
Cited by: 0
References: 4

Abstract

BACKGROUND: Helix-fixation dual-chamber leadless pacemakers reliably provide bradycardia pacing, yet mechanical complexity of the delivery system can rarely lead to procedural failure. CASE PRESENTATION: During atrial leadless pacemaker implantation, resistance felt when transitioning to tether mode suggested tether entrapment within a damaged torque shaft, resulting in mechanical distortion or fracture and preventing both device release and redocking. A controlled bailout maneuver-advancing the protective sleeve near the helix and rotating the entire catheter system counterclockwise-enabled safe unscrewing and withdrawal without complications. CONCLUSION: This report describes a previously unreported tether-related malfunction and a practical maneuver for resolving this failure mechanism.

Fields of Study & Subjects

Cardiac pacing and defibrillation studies Cardiac Arrhythmias and Treatments Cardiac electrophysiology and arrhythmias

Publication details

Journal / source
Journal of Cardiovascular Electrophysiology
Volume
37
Issue
4
Pages
880–883
Publication date
28 January 2026