Tether Entrapment and Fracture Preventing Release and Redocking of a Helix‐Fixation Atrial Leadless Pacemaker: Successful Resolution With a Protective Sleeve–Assisted Bailout Maneuver
Abstract
BACKGROUND: Helix-fixation dual-chamber leadless pacemakers reliably provide bradycardia pacing, yet mechanical complexity of the delivery system can rarely lead to procedural failure. CASE PRESENTATION: During atrial leadless pacemaker implantation, resistance felt when transitioning to tether mode suggested tether entrapment within a damaged torque shaft, resulting in mechanical distortion or fracture and preventing both device release and redocking. A controlled bailout maneuver-advancing the protective sleeve near the helix and rotating the entire catheter system counterclockwise-enabled safe unscrewing and withdrawal without complications. CONCLUSION: This report describes a previously unreported tether-related malfunction and a practical maneuver for resolving this failure mechanism.
Fields of Study & Subjects
Cardiac pacing and defibrillation studies
Cardiac Arrhythmias and Treatments
Cardiac electrophysiology and arrhythmias
Publication details
- Journal / source
- Journal of Cardiovascular Electrophysiology
- Volume
- 37
- Issue
- 4
- Pages
- 880–883
- Publication date
- 28 January 2026
- Language
- EN