Deep learning-based chemiluminescent point-of-care assay for high-sensitivity troponin testing
Abstract
We present a chemiluminescence vertical flow assay (CL-VFA) that achieves high-sensitivity detection of cardiac troponin I (cTnI), a gold standard biomarker for diagnosing myocardial infarction. The CL-VFA system integrates a chemiluminescent paper-based sensor, a portable and low-cost Raspberry Pi-based reader, and a deep learning-powered concentration inference algorithm. This system can rapidly measure cTnI from a 50 µL serum sample within 25 min, reaching a detection limit of 0.16 pg/mL and covering a wide dynamic range of six orders-of-magnitude in concentration. The CL-VFA exceeds the sensitivity of commercial benchtop analyzers by an order-of-magnitude while maintaining a compact and lightweight design. In a blinded clinical study using 66 patient serum samples, the CL-VFA showed a strong correlation with results from an FDA-cleared analyzer, highlighting its reliability and robustness. The platform's affordability, ease-of-use, and high analytical performance make it ideal for use in under-resourced settings, offering a transformative step toward democratizing high-sensitivity diagnostics.
Fields of Study & Subjects
Biosensors and Analytical Detection
Electrochemical sensors and biosensors
Advanced Biosensing Techniques and Applications
Publication details
- Journal / source
- Optical Diagnostics and Sensing: Toward Point-of-Care Diagnostics
- Pages
- 14–14
- Publication date
- 4 March 2026
- Language
- EN