A wearable fluorescence sensor for real-time on-body monitoring of catheter-associated urinary tract infections

View at publisher
Authors: Majed Othman Althumayri, Hatice Ceylan Koydemir, Weiming Xu, Azra Yaprak Tarman
Cited by: 0

Abstract

We present a wearable, fluorescence-based sensor for continuous monitoring of Escherichia coli (E. coli) in catheter bags, designed for the early detection of catheter-associated urinary tract infections (CAUTIs). The system uses a flexible circuit board integrated with UV LEDs and a gray-color sensor to detect fluorescence generated by the endogenous enzymatic activity of E.coli on the MUG substrate. Data is transmitted via Bluetooth to a smartphone application, enabling real-time, cloud-independent analysis. The device detected E. coli concentrations from 100 to 10⁵ CFU/mL within 9 to 3.5 hours, respectively, and demonstrated high specificity against other common pathogens, including Staphylococcus epidermidis, Pseudomonas aeruginosa, and Klebsiella pneumoniae. In vitro bladder model testing further validated its practical use in realistic conditions. At a cost of ~$47 and a weight of only 72 g, this platform provides a cost-effective, portable solution for early CAUTI detection in both clinical and home settings.

Fields of Study & Subjects

Biosensors and Analytical Detection Urinary Tract Infections Management Analytical Chemistry and Sensors

Publication details

Journal / source
Optical Diagnostics and Sensing: Toward Point-of-Care Diagnostics
Pages
2–2
Publication date
4 March 2026
Language
EN