Understanding corneal biomechanical influence in intraocular pressure estimations using optical coherence elastography with a multi-excitation probe

View at publisher
Authors: Etsel Suárez, Alvaro Cigaran, Cristian Bocanegra, Fernando Zvietcovich
Cited by: 0

Abstract

This study investigated the influence of corneal biomechanics on intraocular pressure (IOP) estimation, particularly in normotensive glaucoma patients. Using Phase Sensitive Optical Coherence Tomography with air-puff and ultrasonic systems, we analyzed the correlation between corneal phase speed (Young's modulus), thickness, and controlled IOP (5-25 mmHg) with IOP-related parameters (apex velocity, applanation pressure, displacement) in eye phantoms. Strong correlations were found between phase speed and several IOP parameters (e.g., R = 0.914 for V2), indicating that corneal biomechanics significantly influence IOP estimations. Weaker, but still present, correlations existed between thickness and controlled IOP with these parameters. These findings highlight the importance of considering corneal biomechanics for accurate IOP assessment and glaucoma diagnosis.

Fields of Study & Subjects

Optical Coherence Tomography Applications Corneal surgery and disorders Glaucoma and retinal disorders

Publication details

Journal / source
Optical Elastography and Tissue Biomechanics
Pages
24–24
Publication date
4 March 2026
Language
EN