An improved algorithm for reflective and transmissive pulse oximetry across different skin pigmentation

View at publisher
Authors: Samantha Unger, Rawad Alkallas, Megh Rathod, Jonathan Wu, Daniel Franklin
Cited by: 0
References: 17

Abstract

While pulse oximeters are used ubiquitously in different clinical settings, numerous studies have noted differences in performance across individuals with different skin tones. In a study of 11 participants spanning the Monk Skin Tone scale, we observed a bias of 1.38% and ARMS of 1.31% in transmissive mode and a bias of 3.91% and an ARMS of 2.35% in reflective mode when applying the conventional ratio-of-ratios expression. Adding a red/infrared ratio to the fit reduced the bias and ARMS to -0.17% and 1.11%, respectively in transmissive mode and 0.28% and 1.64%, respectively in reflective mode. This work introduces a readily available optical platform for future investigation and improvement of pulse oximeter accuracy, avoiding the confounder of proprietary signal processing and enabling open science.

Fields of Study & Subjects

Non-Invasive Vital Sign Monitoring Optical Imaging and Spectroscopy Techniques Spectroscopy Techniques in Biomedical and Chemical Research

Publication details

Journal / source
Optical Diagnostics and Sensing: Toward Point-of-Care Diagnostics
Pages
8–8
Publication date
4 March 2026
Language
EN