An empirical model of scatter-induced SDNR degradation in triple-energy photon-counting x-ray imaging

View at publisher Summarise with VIRAT
Authors: Kaitlyn Sims, Jesse Tanguay
Cited by: 0
References: 5

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

Triple-energy photon-counting x-ray imaging produces iodine-specific images by suppressing both soft tissue and bone clutter, offering a potential alternative to dual-energy methods and digital subtraction angiography. A major challenge, however, is scattered radiation, which reduces iodine contrast and reduces signal-difference-to-noise ratio (SDNR). The combined effects of scatter and iodine attenuation are inherently nonlinear. We develop an empirical model of scatter-induced degradation in SDNR for triple-energy imaging. Experiments were performed with a CdTe-based photon-counting detector incorporating analog charge summing. Stacked PMMA phantoms with thicknesses of 10cm, 15cm, and 20cm were imaged at 80kV. The signal slab contained iodine wells with areal densities ranging from 33 to 140mg/cm2, along with aluminum and PMMA perturbations used to calibrate triple-energy subtraction weights. Scatter levels were varied by adjusting the irradiated field-of-view, producing effective scatter fractions (SFeff) from approximately 0.2 to 0.7. Scatter fractions were estimated using the air-gap method for each energy bin and combined into SFeff. The iodine signal difference demonstrated a nonlinear saturation with increasing areal density, most pronounced at higher scatter fractions and for thicker phantoms. SDNR decreased substantially with increasing SFeff, but could not be explained by first-order attenuation alone. After normalization by (1−SFeff)1/2, measured SDNR values were well described by a polynomial empirical model incorporating iodine-dependent terms. The model achieved strong agreement with experimental data across all phantom thicknesses, with R2 values ranging from 0.92 to 0.94. These results demonstrate that scatter and iodine attenuation interact in a nonlinear fashion that strongly influences iodine detectability in triple-energy photon-counting imaging. The empirical model provides a compact description of these effects and offers a framework for protocol optimization.

Fields of Study & Subjects

Advanced X-ray and CT Imaging Digital Radiography and Breast Imaging Cardiac Imaging and Diagnostics

Publication details

Journal / source
SPIE Medical Imaging: Physics of Medical Imaging
Pages
17–17
Publication date
2 April 2026
Language
EN