Study of ion detection efficiency in Jet Counter nanodosimeter

Read full text
Authors: Mateusz Filipek, Aleksandr Bancer, B. Brzozowska, Zygmunt Szefliński, Marcin Pietrzak
Cited by: 0
References: 13

Abstract

Abstract This work reevaluates the ion detection efficiency of the Jet Counter nanodosimeter working with low-pressure nitrogen. SIMION simulations were used to model the trajectories of ions during their extraction from the interaction chamber. The study examined the relationship between efficiency vs. extraction voltage, gas density, and temperature. The highest detection efficiency (approximately 60%) was achieved at a high extraction voltage and low temperature. The findings were then used as a parameter in the Monte Carlo modeling of the Jet Counter operations, and the results were compared with experimental results, showing satisfactory agreement.

Fields of Study & Subjects

Particle Detector Development and Performance Ion-surface interactions and analysis Radiation Detection and Scintillator Technologies

Publication details

Journal / source
EPJ Techniques and Instrumentation
Volume
13
Issue
1
Publication date
9 February 2026