Green Chemistry Spectrofluorimetric Assessment of Folic Acid in Pharmaceutical Formulations Using Acriflavine as an Efficient Fluorescent Probe

View at publisher
Authors: Huda M. Younis, Mohamed Aboel-Kasem Abdel-Lateef, Amal A. Mohamed
Cited by: 0
References: 43

Abstract

A sensitive, simple, inexpensive, rapid, and eco-friendly spectrofluorimetric method was created to assess the folic acid (FA) concentration in tablets using acriflavine (ACF) as an eco-friendly photoprobe. Based on its ability to quench the ACF fluorescence intensity in water at pH 8.0 and λ ex = 460nm. FA concentration was measured by quenching the fluorescence intensity of the ACF at λ em =508 nm within the linear range of 3.5 × 10 –6 –30.0 × 10 –6 mol L –1 with a correlation coefficient r 2 = 0.9991. The limit of quantification (LOQ) and the limit of detection (LOD) are 1.159 × 10 −6 mol·L –1 and 0.383 × 10 −6 mol·L –1 , respectively. This method was easy to use and accurately and effectively evaluated FA in pharmaceutical tablet samples. No influence was observed from the excipients usually contained in pharmaceutical formulations. The method was verified as valid according to the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) guidelines. In addition, its greenness has been estimated using environmental assessment tools.

Fields of Study & Subjects

Analytical Methods in Pharmaceuticals Folate and B Vitamins Research Analytical chemistry methods development

Publication details

Journal / source
Applied Spectroscopy
Volume
80
Issue
5
Pages
446–456
Publication date
16 February 2026