Evaluation of repeatability and reproducibility of central and peripheral axial length measurements in children using the novel Swan optical biometer
Abstract
ObjectiveTo evaluate the repeatability and reproducibility of central axial length and peripheral axial length measurements in children using the Swan 660 M.MethodsThis cross-sectional study was conducted from August to September 2024. A total of 60 children (6-15 years; no significant ocular diseases) were recruited from Zhoupu Hospital, Shanghai (30 males and 30 females; mean age 10.67 ± 2.47 years), and only right eyes were included in the analysis. The Swan 660 M was used to measure central axial length and peripheral axial length. Measurement variability across operator and time was evaluated using the intra-group SD, test-retest repeatability, coefficient of variation, and intraclass correlation coefficient.ResultsThe Swan 660 M demonstrated good repeatability and reproducibility for the measurements of central axial length and most peripheral axial length points, including Nasal 1, Temporal 1, Temporal 2, and Temporal 3 (coefficient of variation 0.9). However, measurement reliability was lower for certain peripheral axial length points, such as Nasal 5 (intraclass correlation coefficient