Performance verification of blood vessel detection system using near-infrared light

View at publisher
Authors: Ji‐Eun Park, Junghun Kim
Cited by: 0
References: 13

Abstract

BackgroundNear-infrared (NIR) imaging technology has been increasingly applied in clinical and biomedical engineering fields for subcutaneous vein detection and vascular guidance. However, most existing systems still face limitations such as inconsistent visualization under various skin tones and difficulties in achieving real-time projection alignment.ObjectiveThe objective of this study is to develop a near-infrared (NIR)-based vascular detection system and to evaluate its performance.MethodsThe developed system provides intuitive visualization by projecting vascular images in real time onto the same skin region. Performance validation was conducted using a vascular model and the abdominal vessels of rabbits.ResultsIn the vascular model experiment, the localization accuracy was 98.9%, and in the rabbit experiment, the detection success rate for vessels with a diameter of ≥1 mm was 100%.ConclusionsThe proposed method was confirmed to reduce the time required for acquiring anatomical knowledge and to minimize complications by preventing incorrect venipuncture.

Fields of Study & Subjects

Optical Imaging and Spectroscopy Techniques Infrared Thermography in Medicine Medical Image Segmentation Techniques

Publication details

Journal / source
Technology and Health Care
Volume
34
Issue
2
Pages
313–321
Publication date
16 February 2026
Language
EN