Ablation catheter entrapment in the mitral valve apparatus during ablation of premature ventricular complexes: a case report

Read full text
Authors: Fan Wang, Xuelian Song, Litian Liu, Feifei Zhang, Xiaoning Li, Yi Dang
Cited by: 0
References: 10

Abstract

Entrapment of the ablation catheter in the mitral valve apparatus during transaortic ablation of premature ventricular complexes is a rare complication that usually requires surgical removal. Random rotating and pulling of the catheter exacerbate the extent of the entrapment, causing chest pain and a drop in blood pressure. The catheter can be loosened by changing the direction of the vector at the tip of the catheter, pulling in the reverse direction and rotating. We describe the case of the thought process and catheter maneuvers used to diagnose, loosen, and remove a trapped catheter.

Fields of Study & Subjects

Cardiac Arrhythmias and Treatments Atrial Fibrillation Management and Outcomes Cardiac Structural Anomalies and Repair

Publication details

Journal / source
Journal of Cardiothoracic Surgery
Volume
21
Issue
1
Pages
68–68
Publication date
4 January 2026
Language
EN