Variation of Magnetic Particle Imaging Tracer Performance with Amplitude and Frequency of the Applied Magnetic Field
Abstract
The magnetic response of magnetic particle imaging (MPI) tracers varies with the slew rate of the applied magnetic field, as well as with the tracer’s average magnetic core size. Currently, 25 kHz and 20 mT/µ 0 drive fields are common in MPI, but lower field amplitudes may be necessary for patient safety in future designs. We studied how several different sizes of monodisperse MPI tracers behaved under different drive field amplitude and frequency, using magnetic particle spectrometry and ac hysteresis for drive field conditions at 16, 26, and 40 kHz, with field amplitudes from 5 to 40 mT/µ 0 . We observed that both field amplitude and frequency can influence the tracer behavior, but that the magnetic behavior is consistent when the slew rate (the product of field amplitude and frequency) is consistent. However, smaller amplitudes provide a correspondingly smaller field of view, sometimes resulting in excitation of a minor hysteresis loop.
Fields of Study & Subjects
Characterization and Applications of Magnetic Nanoparticles
Geomagnetism and Paleomagnetism Studies
Advanced MRI Techniques and Applications
Publication details
- Pages
- 789–794
- Publication date
- 19 January 2026
- Language
- EN
- Document type
- Book chapter