59639 Lighting-Invariant Computational Pupillometry in the Neuro-ICU: Evaluation of the PuRe Pupillometer

View at publisher Summarise with VIRAT
Authors: Patryk Szczęśniewski, Małgorzata Dudzic, Karol Badowski, Michał Świątek, Martyna Płatników, Jakub Klawiter, Marta Kowalow, Wiktor Olszewski, Zuzanna Paryzek, Kacper Pawlak, Hugo Chróst, Michal Wlodarski
+4 more Marek Jacek Dziubinski, Radosław Chrapkiewicz, Artur Drużdż, Bartosz Sokół
Cited by: 0

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

INTRODUCTION: Quantitative pupillometry enhances neurologic assessment but is still hampered by ambient-light sensitivity and limited on-device processing. The PuReTM Pupillometer (Solvemed inc.) off-loads deep-learning algorithms to the neural and graphical processing units of a smartphone, generating a lighting-invariant Pupil Reactivity (PuReTM) score. METHODS: We performed a prospective, single-center observational study in a municipal neuro-ICU. Twelve adults (75% male; median age 59 y) with stroke, traumatic brain injury, hydrocephalus, brain tumor, or intracranial hypertension underwent 1,331 pupillometry recordings (median 44 per patient) under dim (4 lux) to bright (1,200 lux) conditions while receiving standard sedatives/analgesics. The study received institutional ethics committee approval (#419/24) from the Poznan University of Medical Sciences. Frame-wise pupil diameters and PuRe scores were computed on-device; analyses used Python (v3.11). RESULTS: Accuracy: Frame-to-frame diameter error was σ = 0.025 mm across iris colours, surpassing reported infrared-pupillometer precision. Lighting invariance: Conventional PLR parameters differed significantly between dim and bright light (p < 0.001), whereas PuRe scores did not (p = 0.449). Neurologic severity: PuRe was lower during severe neurologic status (GCS ≤ 8; median 1.85 ± 1.39) vs non-severe (median 4.11 ± 0.76; p < 0.001); ROC AUC = 0.94, sensitivity 84.3%, specificity 90.2%. Outcome prediction: Non-survivors showed markedly reduced PuRe (median 0.00 ± 1.15) compared with survivors (median 2.82 ± 1.62; p < 0.001). CONCLUSIONS: The PuRe Pupillometer delivers sub-0.03 mm precision and lighting-invariant reactivity metrics directly on a mobile device. PuRe strongly discriminates neurologic severity and predicts survival, supporting its integration into critical-care workflows as a fast, non-invasive, and resource-efficient neurologic monitor.

Fields of Study & Subjects

Gaze Tracking and Assistive Technology Ophthalmology and Visual Health Research Traumatic Brain Injury Research

Publication details

Journal / source
Neurosurgery
Volume
72
Issue
Supplement_1
Pages
308–308
Publication date
26 March 2026
Language