Dynamic calibration of the air-bearing table for the swing arm profilometer test

View at publisher
Authors: Ling Hua Xiong, Hao Chen, Xiao Luo, Haixiang Hu, zhenyu liu, Qiang Cheng, Feng Zhang, Xuejun Zhang
Cited by: 0
References: 9

Abstract

The swing arm profilometer (SAP) has great potential to achieve high-accuracy measurements for large-aperture optic surfaces. To promote the precision of the SAP further, we proposed a dynamic calibration setup with four high-accuracy capacitive probes and a standard circular ring flat to monitor error motions of the air-bearing table of the SAP during surface measurement. We implemented an experiment on a ∅400 mm flat and achieved an accuracy of 15 nm. Compared with the traditional SAP test results, the accuracy was improved by 25%. Experimenting on a ∅2000 mm off-axis aspheric surface displayed that the ring-shaped system errors from the SAP test were eliminated with a 12% improvement in accuracy. It exhibited a good suppression effect on rotationally symmetric errors from the SAP.

Fields of Study & Subjects

Advanced Measurement and Metrology Techniques Optical measurement and interference techniques Surface Roughness and Optical Measurements

Publication details

Journal / source
Applied Optics
Volume
65
Issue
8
Pages
2635–2635
Publication date
13 February 2026
Language