Simple calibration procedure for a full-Stokes imaging polarimeter using a polarized camera
Abstract
A simple calibration procedure and data-extraction process are presented for a full-Stokes imaging polarimeter using a polarized camera and a liquid-crystal variable retarder (LCVR). The LCVR is calibrated using combinations of measured polarization images with a horizontal linear polarizer as a sample. Then, the Stokes vector is measured using two images with two values of the LCVR retardance. The proposed calibration procedure is simple and robust and can be useful for real-time calibration in the field.
Fields of Study & Subjects
Optical Polarization and Ellipsometry
Metamaterials and Metasurfaces Applications
Molecular spectroscopy and chirality
Publication details
- Journal / source
- Applied Optics
- Volume
- 65
- Issue
- 9
- Pages
- 3130–3130
- Publication date
- 27 February 2026
- Language
- EN