Phase diversity improves retinal image quality in adaptive optics scanning light ophthalmoscopy
Abstract
The quality of retinal images is compromised by aberrations that remain uncorrected even in confocal adaptive optics imaging. This study demonstrates phase diversity (PD), a computational imaging technique, to address residual aberrations and enhance image quality in adaptive optics scanning laser ophthalmoscopy (AOSLO). By using images of the same object obtained with and without deliberately added aberrations, PD computes and compensates for the effects of existing residual aberrations beyond those corrected by a closed-loop AO system. Experimental validation demonstrates that PD improves visualization of retinal microstructures, including cone mosaics and dendrites of fluorescently labeled retinal ganglion cells (RGCs).
Fields of Study & Subjects
Ophthalmology and Visual Impairment Studies
Optical Coherence Tomography Applications
Digital Holography and Microscopy
Publication details
- Journal / source
- Biomedical Optics Express
- Publisher
- Optica Publishing Group
- Volume
- 17
- Issue
- 4
- Pages
- 1767–1767
- Publication date
- 4 March 2026
- Language