Neckline-free open-source cap design and improved motion resistance for perioperative fNIRS procedures

Read full text
Authors: Seth Crawford, Shimra Fine, Kyla Covato, Caroline Caveness, Audrey K. Bowden
Cited by: 1
References: 43

Abstract

Functional near-infrared spectroscopy (fNIRS) caps are limited by motion artifacts and unstable optode–scalp coupling, which degrade signal quality during extended and naturalistic monitoring sessions. We present a fabric-based, chinstrap-free headcap designed to improve mechanical stability and optical signal fidelity through enhanced coupling and reduced relative motion. The open-source cap incorporates a two-layer design with slip-resistant materials and a circumferential tightening mechanism. Performance was evaluated over a two-hour protocol in a mock operating room environment, with motion quantified via image-based tracking, signal quality assessed using signal-to-noise ratio (SNR), scalp-coupling index (SCI), and breath-holding responses, and user comfort evaluated with a human factors questionnaire. Results show reduced displacement and robust hemodynamic signals, supporting improved reliability of fNIRS measurements in perioperative conditions.

Fields of Study & Subjects

Intraoperative Neuromonitoring and Anesthetic Effects Thyroid and Parathyroid Surgery Advanced MRI Techniques and Applications

Publication details

Journal / source
Biomedical Optics Express
Publisher
Optica Publishing Group
Volume
17
Issue
10
Pages
5077–5077
Publication date
28 August 2026