Broad linewidth sources result in a skin-tone bias in non-invasive optical measurement of oxygen saturation

Read full text Summarise with VIRAT
Authors: Giles P. Blaney, Sergio Fantini
Cited by: 0

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

Clinical evidence suggests that pulse oximeters overestimate arterial oxygen saturation in patients with darker skin tones. This letter addresses whether the emission of broad linewidth Light Emitting Diodes (LEDs) could explain this melanin dependent bias. Using a hybrid diffusion theory and Monte Carlo model, we simulated pulse oximetry measurements across varying melanin concentrations and source bandwidths. Results indicate that calibration curves depend on melanin when using broad linewidth LEDs but do not depend on melanin for monochromatic Laser Diodes (LDs). Therefore, adopting LD based oximeters could resolve this bias in agreement with conclusions of previous experimental work.

Fields of Study & Subjects

Non-Invasive Vital Sign Monitoring Optical Imaging and Spectroscopy Techniques Analytical Chemistry and Sensors

Publication details

Publication date
28 March 2026
Language
EN
Document type
Preprint