Evaluation of the MAGLUMI non-competitive angiotensin I chemiluminescence immunoassay for measuring renin activity in human plasma
Abstract
OBJECTIVES: Plasma renin activity (PRA) measures the rate of renin converting angiotensinogen to angiotensin I in circulation and reflects the function of renin-angiotensin-aldosterone system (RAAS). The study aimed to demonstrate the performance of the MAGLUMI Angiotensin I assay for the measurement of PRA. METHODS: The MAGLUMI Angiotensin I assay was evaluated for precision, linearity, limit of blank (LoB), limit of detection (LoD), limit of quantification (LoQ), interferences, cross-reactivity and reference intervals. It was compared to liquid chromatography tandem mass spectrometry (LC-MS/MS) using 54 samples, and impact of clinical factors was assessed in 181 samples. RESULTS: The assay demonstrated repeatability of 1.5 %, 1.85%, and 1.6 % across three different pool concentrations, with corresponding within-laboratory imprecision values of 2.0 %, 2.5%, and 1.9 %. The linear range was [0.10, 24.0] ng/mL. The LoB, the LoD, and the LoQ were 0.03 ng/mL, 0.07 ng/mL, and 0.10 ng/mL, respectively. The interferences and cross-reactivity met the criterion for acceptable cross-reactivity (CR) with
Fields of Study & Subjects
Renin-Angiotensin System Studies
Hormonal Regulation and Hypertension
Advanced Proteomics Techniques and Applications
Publication details
- Journal / source
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Publisher
- Walter de Gruyter GmbH
- Volume
- 64
- Issue
- 5
- Pages
- 1142–1151
- Publication date
- 6 January 2026
- Language
- EN