Validation study of the Helena V8 UltraCE capillary elecrophoresis analyser
Abstract
OBJECTIVES: Serum protein electrophoresis is one of the keystone investigations for screening for monoclonal proteins and assessing the serum protein profile. The newly released Helena V8ultra capillary electrophoresis system has been evaluated. METHODS: In total, 164 serum samples were assessed on the Helena V8 UltraCE system and compared with the Sebia Capillarys™ instrument. Abnormalities suggestive of monoclonal proteins were confirmed by immunosubtraction. Imprecision studies intervals were determined. Special attention was paid to complement C3 polymorphism. RESULTS: globulins; -2.317 % for beta globulins; 0.066 % for gamma globulins. Among the 6 samples with monoclonal proteins confirmed by immunofixation, all were seen on both methods, with only 1 discordant result. The theoretical plate number (albumin fraction) was 2,371 ± 718 (Helena V8UltraCE) vs. 445 ± 115 (Sebia Capillarys 2). The high resolution also allows to distinguish Complement C3 phenotypes. CV values for the various fractions were low, ranging from 0.47 % (albumin) to 0.89 % (gamma globulins). IgG, IgA and IgM M-proteins could be detected in the electropherogram with an excellent sensitivity (
Fields of Study & Subjects
Advanced Proteomics Techniques and Applications
Microfluidic and Capillary Electrophoresis Applications
Clinical Laboratory Practices and Quality Control
Publication details
- Journal / source
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Publisher
- Walter de Gruyter GmbH
- Volume
- 64
- Issue
- 6
- Pages
- 1399–1403
- Publication date
- 2 January 2026
- Language
- EN