Comparison of three different instruments for vault measurements after implantable collamer lens implantation
Abstract
AIM: To evaluate the differences and consistency of vault measurements obtained by Scheimpflug tomography (Pentacam), anterior segment optical coherence tomography (AS-OCT, CASIA II), and ultrasound biomicroscopy (UBM) following implantable collamer lens (ICL) V4c implantation. METHODS: Vault measurements were acquired using three modalities: Pentacam, CASIA II AS-OCT, and UBM. Repeated-measures analysis of variance was used to compare the vault values obtained by the three devices. The correlation and consistency of measurements among the three instruments were assessed using the Pearson correlation coefficient, intraclass correlation coefficient (ICC), and Bland-Altman plots. RESULTS: This retrospective study enrolled 210 myopic eyes of 210 patients (158 women and 52 men) who underwent ICL implantation: 108 eyes had a myopic ICL V4c implanted, and 102 eyes had a toric ICL V4c implanted. The mean vault values measured by Pentacam, CASIA II, and UBM were 452.64±204.20 μm, 538.57±203.54 μm, and 560.95±227.54 μm, respectively, with statistically significant differences among the three groups (P
Fields of Study & Subjects
Corneal surgery and disorders
Ophthalmology and Visual Impairment Studies
Glaucoma and retinal disorders
Publication details
- Journal / source
- International Journal of Ophthalmology
- Publisher
- Press of International Journal of Ophthalmology (IJO Press)
- Volume
- 19
- Issue
- 3
- Pages
- 475–482
- Publication date
- 12 February 2026
- Language
- EN