High-precision alignment of an upgraded soft X-ray polarimeter at Diamond Light Source

Read full text Summarise with VIRAT
Authors: Qingxin Meng, Hongchang Wang, Garvey,Ben, Mark Hooper, Adam Howell, Ken Jones, Larkin,Paul, Arindam Majhi, Andrew Malandain, Andrew Peach, Xuan Nam Tran, Guy Wilkin
+2 more Douglas Winter, Kawal J. S. Sawhney
Cited by: 0

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

Knowledge of X-ray beam polarisation on a synchrotron beamline is essential, not only for characterising the undulator performance, but also for precise analysis of dichroic and chiral experiments. The upgraded high-precision soft X-ray polarimeter at Diamond Light Source features multiple retarder adjusters to allow precise concentricity and angular alignment to the analyser. An offline alignment procedure has been developed, achieving 69 µm horizontal and 17 µm vertical concentricity alignment, as well as 4 µrad yaw and 9 µrad pitch alignment. Compared to the original version of the instrument, the vertical concentricity alignment improved by 14× and the pitch alignment improved by 18× $*$. The procedure uses a laser diode to mimic the X-ray beam and a hexapod to align the analyser. Concentricity alignment relies on monitoring the intensity as the laser beam is cropped by plane mirrors on the retarder and analyser stages. Angular alignment is achieved by measuring the retarder and analyser rotation vectors using an autocollimator. The improved alignment allows the polarimeter to meet the stringent requirements for complete polarisation measurement above 1 keV.

Fields of Study & Subjects

Advanced X-ray Imaging Techniques Particle Accelerators and Free-Electron Lasers X-ray Spectroscopy and Fluorescence Analysis

Publication details

Journal / source
JACOW
Publication date
2 April 2026
Language
EN