Optimizing SGM beamline performance: hexapod and spectroscopy enhancements

Read full text Summarise with VIRAT
Authors: Tor Pedersen, James Jay Dynes, Sandra LeBlanc, Tom Regier
Cited by: 0

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

Recent developments on the Spherical Grating Monochromator (SGM) beamline at the Canadian Light Source (CLS) have significantly enhanced its capabilities, particularly through the integration of a vacuum-compatible Physik Instrumente hexapod (H-811.I2V) and the implementation of Bluesky data acquisition software. These upgrades have facilitated the transition from traditional X-ray Absorption Spectroscopy (XAS) measurements to advanced spectromicroscopy techniques. The hexapod allows for sub-micron scale sample manipulation, enabling high-resolution imaging with a 20 mm × 15 mm field of view. Additionally, the modelling of the Kirkpatrick-Baez (KB) mirror system for adaptive focusing has further optimized the beamline's performance providing a beam spot size of less than 10 µm². These developments have not only significantly improved the beamline's capabilities for environmental and catalytic material studies, but also increased the data quality for all routine spectroscopy measurements conducted on the beamline.

Fields of Study & Subjects

Advanced X-ray Imaging Techniques X-ray Spectroscopy and Fluorescence Analysis Particle Accelerators and Free-Electron Lasers

Publication details

Journal / source
JACOW
Publication date
2 April 2026
Language
EN